Choosing the Correct Test Target

Test targets are useful when evaluating or calibrating an imaging system's performance or image quality. This could include troubleshooting the system, certifying or evaluating measurements, as well as establishing a foundation to ensure the system works well with another. Image quality can be defined by different components, particularly resolution, contrast, modulation transfer function (MTF), depth of field (DOF), and distortion; therefore, one or more types of test targets may be necessary or helpful depending upon the type of system being constructed or what needs to be measured. Fortunay, an array of targets exists that cater towards specific systems including cameras, visual displays, or even a single, thin lens. To be able to choose the correct test target, it is important to first understand the components of image quality.

 

COMPONENTS OF IMAGE QUALITY

Resolution

Resolution is an imaging system's ability to distinguish object detail. It is often expressed in terms of line-pairs per millimeter (lp/mm) as seen in Figure 1. A low resolution image usually lacks fine detail and is often blurry, whereas a high resolution image is highly detailed and clear.

Figure 1: Relation of Line-Pairs to Square Waves

 

To illustrate this concept, imagine two squares being imaged onto pixels of a CCD camera. Assuming that the primary magnification of the lens is such that one square fills one camera pixel (Figure 2a), if there is no space between the filled pixels, it will appear as one large, red rectangle. However, if "white space", or space distinctively different from the original pixel color, is found between the pixels, the camera will be able to distinguish among the two squares (Figure 2b). Hence, the pairing of the red square and "white space" becomes one lp/mm, which corresponds to two separate pixels.

Figure 2: Pair of Red Squares Unresolved (a) vs. Resolved (b)

 

Contrast

Contrast is a measurement of the separation between the light and dark regions of an image. More specifically, contrast is a change in the intensity or brightness from one point to another. It affects how effectively the differences between the object and the shades of gray in the background are reproduced. An image with the highest contrast is one in which black is truly black and white is truly white, without any shades of gray in between. As contrast is reduced, the distinction between black and white begins to blur, in a very literal sense, and shades of gray appear (Figure 3).

Figure 3: Contrast in Relation to Pixels

Contrast is often expressed in terms of percentage (%) and is calculated by using Maximum Intensity (Imax) and Minimum Intensity (Imin), as expressed in Equation 1. It can also be represented by a periodic function (i.e. square wave or sine wave), or a function that alternates regularly and instantaneously between two levels.

 

Modulation Transfer Function (MTF)

Modulation Transfer Function, or MTF, is a measurement of an imaging lens' ability to transfer contrast from the object plane to the image plane at a specific resolution. The object and image planes are the spatial areas where the object and the image preside. The object plane is in front of the imaging system, and the image plane is either in front or behind the imaging system depending on whether the image is real or virtual. MTF is expressed with respect to image resolution (lp/mm) and contrast (%), as seen in Figure 4. Typically, as resolution increases, contrast decreases until a cut-off point, at which the image becomes irresolvable and grey.

Figure 4: Example MTF Curve of 0.13X PMAG Imaging Lens

 

Another component of MTF, in addition to the aforementioned resolution and contrast, is diffraction limit. Diffraction limit is a physical limit restricting a lens from being able to image points or edges perfectly. Since it is constrained by the wave nature of light, even a "perfectly" designed and manufactured lens cannot achieve diffraction limited performance. However, designers utilize a variety of methods to reduce aberrations and increase overall system accuracy in order to come as close as possible to reaching a system's ideal diffraction limit.

Correspondingly, a len's geometry contributes to its ability to reproduce good quality image. Lens Diameter (D), Focal Length (f) and f/# (Equation 2) all affect MTF.

f/# is the light gathering ability of a lens. As Lens Diameter increases, f/# decreases. Low f/# lenses collect the most light, thereby making them ideal for light restrictive applications. Although high f/# can improve an imaging lens' performance, increasing it too much can be detrimental because it can cause the diffraction limit to become progressively worse.

 

Depth of Field (DOF)

Depth of Field, or DOF, is the ability of a lens to maintain a desired amount of image quality as the object being viewed moves in and out of focus. DOF is defined with an associated resolution and contrast, as both suffer when an object is placed closer or farther from the optimal working distance (Figure 5). DOF also applies to objects with depth, since high DOF lenses can image the whole object clearly. DOF is estimated by a single value calculated from the diffraction limit, making it a theoretical approximation. It is difficult, however, to make a genuine comparison because many imaging lenses are not diffraction limited. For example, two lenses may have the same f/# (i.e. equal diffraction limit), but do not necessarily have similar performance or comparable DOF. Therefore, the only way to truly determine DOF is to use a test target to test it.

Figure 5: Geometric Representation of DOF for Low and High f/# Imaging Lenses

 

Distortion

Distortion is a type of geometrical aberration that causes a difference in magnification of the object at different points in the image. As light rays carry the image of an object through a system, various points are misplaced relative to the center of the field, or the central point of the image. Therefore, distortion is not an aberration that causes blurs, but an aberration that causes a dislocation. Distortion is calculated by:

 

where AD is the Actual Distance and PD is the Paraxial, or Predicted, Distance.

Distortion, represented by a percentage, can be either positive or negative. A positive percentage represents "pincushion" distortion, whereas a negative percentage represents "barrel" distortion. Figure 6 illustrates barrel and pincushion distortion compared to an ideal, perfectly square non-distorted image.

Figure 6: Barrel and Pincushion Distortion

 

Although distortion is present in almost all lenses, it can be corrected by using short focal lengths. Unfortunay, short focal length systems tend to suffer from more diffraction effects than their longer focal length counterparts. Correcting for one component of image quality undeniably affects another, a fact that must always be taken into account. An important concept to keep in mind is that distortion causes information about the object to be misplaced but not lost. The original, un-distorted image can be reconstructed through image analysis software.

 

TYPES OF TEST TARGETS

A variety of test targets exist to help characterize resolution, contrast, modulation transfer function (MTF), depth of field (DOF), and distortion in an imaging system. Use the following guide of the most popular targets to make choosing the correct test target as easy as possible.

 

1951 USAF Resolution Targets

Consist of horizontal and vertical bars organized in groups and elements. Each group is comprised of up to nine elements within a range of twelve groups. Every element is composed of three horizontal and three vertical bars equally spaced with one another within a group and corresponds to an associated resolution based on bar width and space. The vertical bars are used to calculate horizontal resolution and horizontal bars are used to calculate vertical resolution. These targets are very popular when considering a target for testing resolution.

 

Typical Applications

Testing Resolution in Applications such as Optical Test Equipment, Microscopes, High Magnification Video Lenses, Fluorescence and Confocal Microscopy, Photolithography, and Nanotechnology

 

IEEE Resolution Targets

Designed to characterize the amount of resolution a camera or display system is able to reproduce from an original image. Because resolution can be different throughout the field of view, both horizontal and vertical resolution can be measured in the center of the target as well as the four corners. IEEE Resolution Targets can also be used to check scanning, linearity, aspect, shading, and interlacing, as well as measure TV lines.

 

Typical Applications

Testing of Analog Imaging Systems

 

Ronchi Rulings

Consist of a square wave optic with constant bar and space patterns that carry a high contrast ratio. They are ideal for reticle and field calibration requirements, and often used for evaluating resolution, field distortion and parafocal stability. Ronchi Rulings are not limited to only calculating resolution; they can be used for diffraction testing.

 

Typical Applications

Testing the Parameters of Resolution and Contrast, Diffraction Testing

 

Distortion Targets

Used for calibrating imaging systems for distortion, which is a geometrical aberration that may misplace certain parts of the image. These targets consist of a grid of dots that are separated by various distances depending on the application.

 

Typical Applications

Lower Focal Length Lenses, Systems that Carry a Wide Field of View

 

Depths of Field Targets

Depths of Field Targets Directly test the depth of field in imaging systems without the use of calculations. The target should be mounted 45° from the face of the lens that is parallel to the object to be viewed; the scale of the target consists of horizontal and vertical lines that measure frequency in line pairs per mm (lp/mm).

 

Typical Applications

Circuit Board Inspection, Security Cameras

 

Star Targets

Star Targets Ideal for identifying focus errors, astigmatism, as well as other focus differing aberrations. The target consists of a circle formed with alternating black and white radial lines emanating from a central point. Because the lines taper, a continuous change in resolution is present and can be measured in both vertical and horizontal directions without repositioning.

 

Typical Applications

Alignment of a System, Assistance with Assembly, Comparing Highly Resolved or Magnified Imaging Systems

 

EIA Grayscale Targets

EIA Grayscale Targets Useful for testing optical and video inspection systems, and consist of a standard pattern and carry two scales, one linear and the other logarithmic, which is useful depending on the linearity of the detector being used. Each scale has nine steps that are acuy tuned for a precise halftone pattern.

 

Typical Applications

Optical and Video Inspection Systems, Evaluating Contrast Levels in Cameras

 

Color Checker Targets

Color Checker Targets Used to determine true color balance or optical density of any color rendition system. They may be expanded to include more squares with a different assortment of colors and act as a reference for testing and standardizing color inspection and analysis systems.

 

Typical Applications

Color Rendition Systems, Digital Cameras and Photography

 

Exploring 1951 USAF Resolution Targets

1951 USAF Resolution Targets have been and are currently a standard when considering a target that tests the resolution of an imaging system. They consist of horizontal and vertical bars organized in groups and elements. Each group is composed of six elements, and each element is composed of three horizontal and three vertical bars equally spaced with one another. There can be a total of twelve groups, with larger numbers used for higher resolution. For example, a standard resolution 1951 target consists of group numbers from -2 to 7, whereas a high resolution of -2 to 9; the element number is the same. The resolution is based on bar width and space, where the length of the bars is equal to five times the width of a bar (Figure 7). One Line Pair (lp) is equivalent to one black bar and one white bar. Vertical bars are used to calculate horizontal resolution and horizontal bars are used to calculate vertical resolution.

Figure 7: 1951 USAF Target Specifications

 

Qualitatively, the resolution of an imaging system is defined as the group and element combination directly before the black and white bars begin to blur together. Quantitatively, resolution (in terms of line pairs per millimeter, or lp/mm) can be calculated by:

 

It is important to keep in mind that calculating resolution with a 1951 USAF Target is subjective. In other words, it depends on who is looking at the target. Someone with 20/20 vision (using the Snellen Ratio) is able to discern higher resolution than someone with, for example, 20/25 or 20/30 vision. Even though the actual test yields precise resolution values, the user's vision can lead to imprecise measurements.

 

APPLICATION EXAMPLES

Example 1: Calculating Resolution with a 1951 USAF Resolution Target

When given a specified group and element number, one can easily calculate the resolution in lp/mm using Equation 4. For instance, if the vertical or horizontal bars start to blur at group 4 element 3, the resolution of the system can be designated as group 4 element 2. To quickly calculate resolution, use our 1951 USAF Resolution EO Tech Tool.

To convert lp/mm to microns (μm), simply take the reciprocal of the lp/mm resolution value and multiply by 1000.

 

Example 2: Understanding f/#

To understand the relationship between f/#, depth of field, and resolution, consider an example with a 35mm Double Gauss Imaging Lens (Figure 8). In this example, the lens will be integrated into a system which requires a minimum of 5 lp/mm (200μm) object resolution at 20% contrast. The diffraction limit , or cutoff frequency, is determined by Equation 7:

Figure 8: Graphical Representation of Resolution vs. f/# (Left) and DOF vs. f/# (Right) for a 35mm Double Gauss Imaging Lens

 

where λ is the wavelength of the system. For simplicity, Equation 7 assumes a non-aberrated, ideal system. Because this system is expected to have aberrations, though, the diffraction limit decreases with increasing f/#. Determining an ideal f/# for this system leads to calculating the highest possible depth of field. Comparing resolution vs. f/#, it is evident that below f/3, the lens is limited by aberrations and cannot obtain the minimum desired resolution. However, "sping down" or closing the iris reduces aberrations and improves DOF. At f/4.2, diffraction effects caused by the optical elements within the imaging lens become more prominent than the effects from aberrations; this is the point at which the lens becomes diffraction limited. Beyond f/4.2, closing the aperture increases DOF, but reduces resolution. At f/13.5, the diffraction limit defines the extent of the desired resolution. Beyond f/13.5, resolution continues to decrease while DOF continues to increase. In this particular example, f/13.5 is the ideal f/# for an optimum depth of field at a minimum resolution.

聯系我們

地址:江蘇省江陰市人民東路1091號1017室 傳真:0510-68836817 Email:sales@rympo.com
24小時在線客服,為您服務!

版權所有 © 2025 江陰韻翔光電技術有限公司 備案號:蘇ICP備16003332號-1 技術支持:化工儀器網 管理登陸 GoogleSitemap

在線咨詢
QQ客服
QQ:17041053
電話咨詢
0510-68836815
關注微信
欧美成人午夜激情视频_亚洲视频第一页_日韩免费看的电影电视剧大全_欧美黄色片视频_欧美性色视频在线_亚洲第一福利视频_欧美亚洲激情视频_成人欧美在线视频_亚洲综合一区二区不卡_国产91精品久久久久久久_成人午夜两性视频_日韩欧美在线视频免费观看_久久人人爽人人_国产成人综合久久_国产精品久久久久9999_久久99热精品这里久久精品
97精品免费视频| 国产精品91在线观看| 国产欧美日韩精品专区| 国产一区二区三区在线视频| 中文字幕在线看视频国产欧美在线看完整| 亚洲bt欧美bt日本bt| 久久影院资源网| 欧美巨猛xxxx猛交黑人97人| 97超视频免费观看| 亚洲影院污污.| 91网站在线看| 51久久精品夜色国产麻豆| 亚洲最大的免费| 欧美激情按摩在线| 欧美午夜www高清视频| 日韩中文视频免费在线观看| 91精品国产高清久久久久久久久| 亚洲欧美日韩一区二区三区在线| 亚洲第一视频网| 欧美肥臀大乳一区二区免费视频| 日韩女优人人人人射在线视频| 国产一级揄自揄精品视频| 亚洲丝袜av一区| 欧美极品少妇xxxxⅹ裸体艺术| 久久久久久av| 色综合91久久精品中文字幕| 精品亚洲一区二区三区在线观看| 懂色av一区二区三区| 日韩在线视频免费观看| 亚洲免费人成在线视频观看| 一区三区二区视频| 亚洲精品天天看| 欧美午夜电影在线| 国产视频精品va久久久久久| 欧美刺激性大交免费视频| 欧美福利在线观看| 欧美在线xxx| 色先锋久久影院av| 精品国产电影一区| 中日韩美女免费视频网站在线观看| 日韩欧美在线网址| 成人深夜直播免费观看| 国产成人精品最新| 国产精品久久9| 亚洲第一区第一页| 国产97免费视| 在线电影中文日韩| 成人午夜两性视频| xxav国产精品美女主播| 国产午夜精品理论片a级探花| 国内精品在线一区| 久久精品视频中文字幕| 久久久久久久999精品视频| 96精品久久久久中文字幕| 亚洲片国产一区一级在线观看| 亚洲一区二区精品| 亚洲美女精品成人在线视频| 中文字幕精品www乱入免费视频| 亚洲综合在线做性| 精品丝袜一区二区三区| 精品久久久中文| 亚洲精品一区在线观看香蕉| 色婷婷av一区二区三区久久| 日韩黄色高清视频| 色婷婷av一区二区三区在线观看| 91亚洲国产精品| 久久精品99国产精品酒店日本| 欧美怡春院一区二区三区| 亚洲国产91精品在线观看| 91超碰caoporn97人人| 欧美精品制服第一页| 91精品久久久久久久久久另类| 亚洲电影免费观看高清完整版| 日韩成人高清在线| 欧美丰满少妇xxxxx做受| 亚洲欧美中文日韩v在线观看| 国产欧美日韩精品专区| 亚洲综合最新在线| 欧洲亚洲免费在线| 亚洲**2019国产| 国产一区二区三区欧美| 日韩二区三区在线| 另类视频在线观看| 亚洲精品久久7777777| 亚洲综合中文字幕在线| 欧美亚洲午夜视频在线观看| 日韩欧美成人免费视频| 国产精品久久久久久久9999| 亚洲黄色成人网| 亚洲成av人乱码色午夜| 亚洲欧美制服第一页| 精品国内产的精品视频在线观看| 亚洲精品电影网在线观看| 日韩电影中文字幕在线观看| 欧美黑人极品猛少妇色xxxxx| 亚洲欧美中文字幕在线一区| 精品国产1区2区| 国产精品久久久久999| 亚洲新中文字幕| 欧美激情视频播放| 亚洲va久久久噜噜噜| 一个人看的www久久| 欧美一级淫片aaaaaaa视频| 91国产高清在线| 欧美另类极品videosbest最新版本| 欧美性生交xxxxx久久久| 91av中文字幕| 91香蕉嫩草神马影院在线观看| 亚洲最大成人免费视频| 九九久久国产精品| 国外成人在线视频| 91国内在线视频| 久久久久久一区二区三区| 国产一区红桃视频| 国产精品一区二区三| 国产乱肥老妇国产一区二| 永久免费看mv网站入口亚洲| 欧美大荫蒂xxx| 日本道色综合久久影院| 久久久久久久久久国产| 日韩av大片免费看| 97精品免费视频| 亚洲国产精品久久久久| 永久555www成人免费| 久久久久久久久久久国产| 欧美性猛交99久久久久99按摩| 亚洲偷熟乱区亚洲香蕉av| 欧美野外猛男的大粗鳮| 96精品视频在线| 91精品中文在线| 亚洲自拍欧美色图| 亚洲bt欧美bt日本bt| 日韩av免费在线播放| 欧美亚洲国产精品| 日本午夜精品理论片a级appf发布| 午夜精品久久久久久99热| 亚洲aaa激情| 欧美大全免费观看电视剧大泉洋| 精品国产自在精品国产浪潮| 中文字幕亚洲天堂| 国产精品久久久久久亚洲影视| 日本成熟性欧美| 久久久久久久国产精品视频| 最近中文字幕mv在线一区二区三区四区| 亚洲国产精品va在看黑人| 66m—66摸成人免费视频| 欧美自拍视频在线观看| 日韩网站免费观看| 成人午夜激情网| 91av免费观看91av精品在线| 欧美一级淫片丝袜脚交| 韩日欧美一区二区| 亚洲天堂av电影| 亚洲第一中文字幕| 日本中文字幕久久看| 91精品国产综合久久男男| 国产精品h在线观看| 亚洲欧美综合区自拍另类| 免费成人高清视频| 中文字幕亚洲激情| 欧美一级淫片播放口| 中文字幕欧美专区| 亚洲午夜精品视频| 九九久久久久久久久激情|